基于PANA 9600E全自动核酸工作站的超敏HBV DNA定量检测性能评价

徐远东, 耿帜, 包一熙, 等. 基于PANA 9600E全自动核酸工作站的超敏HBV DNA定量检测性能评价[J]. 临床血液学杂志, 2022, 35(4): 239-243. doi: 10.13201/j.issn.1004-2806.2022.04.001
引用本文: 徐远东, 耿帜, 包一熙, 等. 基于PANA 9600E全自动核酸工作站的超敏HBV DNA定量检测性能评价[J]. 临床血液学杂志, 2022, 35(4): 239-243. doi: 10.13201/j.issn.1004-2806.2022.04.001
XU Yuandong, GENG Zhi, BAO Yixi, et al. Performance evaluation of PANA 9600E automatic nucleic acid extraction platform for ultra-sensitive HBV DNA quantitative detection[J]. J Clin Hematol, 2022, 35(4): 239-243. doi: 10.13201/j.issn.1004-2806.2022.04.001
Citation: XU Yuandong, GENG Zhi, BAO Yixi, et al. Performance evaluation of PANA 9600E automatic nucleic acid extraction platform for ultra-sensitive HBV DNA quantitative detection[J]. J Clin Hematol, 2022, 35(4): 239-243. doi: 10.13201/j.issn.1004-2806.2022.04.001

基于PANA 9600E全自动核酸工作站的超敏HBV DNA定量检测性能评价

详细信息

Performance evaluation of PANA 9600E automatic nucleic acid extraction platform for ultra-sensitive HBV DNA quantitative detection

More Information
  • 目的 对全自动核酸工作站PANA 9600E的超敏HBV DNA定量检测性能进行评价, 评估其是否满足临床应用要求。方法 参考美国临床和实验室标准化协会(CLSI)和中国合格评定国家认可委员会(CNAS)等相关文件, 采用天隆PANA 9600E全自动核酸工作站及其配套HBV DNA超敏定量检测试剂和GENTIER 96E实时荧光定量PCR仪, 应用HBV DNA血清标准物质、临床混合血清样本和乙型肝炎病毒核酸国家标准品, 对该检测系统的正确度、精密度、线性范围、定量限、检出限、防交叉污染、抗干扰能力等性能进行评价;同时与NP968-C半自动系统检测56例临床血清样本进行方法学比对。结果 ① PANA 9600E系统检测HBV DNA血清标准物3个浓度4.60×106、1.41×103和2.00×102 IU/mL的均值与靶值的差值绝对值均小于0.40 log10IU/mL, 正确度满足行业标准;检测2.0×102、2.0×103、2.0×104和2.0×106 IU/mL 4个浓度的临床混合血清样本, 重复性精密度CV分别为3.93%、1.50%、0.89%和0.54%, 中间精密度CV分别为5.00%、1.95%、1.36%和1.35%, 精密度验证通过;②PANA 9600E系统在3.0×101IU/mL~3.0×108IU/mL内呈线性, 线性回归方程为Y=1.028 5X-0.075 8, R2=0.999 4;定量限为30 IU/mL, 检出限为10 IU/mL;防交叉污染验证通过, 阴性样本检测结果均为阴性;当胆红素≤512 μmol/L、血红蛋白≤100 mg/mL、甘油三酯≤18 mmol/L时对该系统的检测结果无明显影响, 抗干扰能力验证通过;③对于56例临床血清样本, PANA 9600E和NP968-C系统检出的阳性率分别为60.71%(34/56)和57.14%(32/56), 差异无统计学意义(P>0.05), Kappa值为0.852 6(95%CI: 0.713 9~0.991 4), 一致性较好;对于定量范围内的19例临床血清样本, PANA 9600E和NP968-C系统的检测值分别为3.954±1.850、3.828±1.916 log10IU/mL, 差异有统计学意义(P< 0.05);进一步进行Pearson相关和Bland-Altman图分析, 回归方程为Y=1.030 9X-0.249 5, R2=0.990 5;2种方法检测结果差值为(0.125 8±0.195 3) log10IU/mL, 95%一致性界限为-0.256 9~0.508 5 log10IU/mL, 94.74%(18/19)的点在95%一致性界限以内, 界外点数为1(5.26%)。结论 基于PANA 9600E全自动核酸工作站的超敏HBV DNA定量检测系统的正确度、精密度、线性范围、定量限、检出限、抗交叉污染、抗干扰能力等性能良好;定量限为30 IU/mL, 检出限为10 IU/mL, 达到超敏检测的要求;与现用的NP968-C半自动系统的检测结果具有可比性, 相关性和一致性好, 可满足临床应用要求。
  • 加载中
  • 图 1  PANA 9600E检测系统线性范围验证结果

    图 2  2种方法定量检测HBV DNA结果的相关性分析

    图 3  2种方法定量检测HBV DNA结果的Bland-Altman图分析

    表 1  PANA 9600E检测系统正确度验证结果 log10IU/mL

    HBV DNA国家标准物质 检测均值 靶值 差值
    GBW(E)090139(S2) 6.34 6.66 -0.32
    GBW(E)090137(S5) 3.13 3.15 -0.02
    GBW(E)090586(S7) 2.01 2.30 -0.29
    下载: 导出CSV

    表 2  PANA 9600E检测系统精密度验证结果 log10IU/mL

    浓度(IU/mL) 重复性精密度 中间精密度
    X±S CV/% X±S CV/%
    2.0×102 2.27±0.09 3.93 2.18±0.11 5.00
    2.0×103 3.21±0.05 1.50 3.19±0.06 1.95
    2.0×104 4.16±0.04 0.89 4.23±0.06 1.36
    2.0×106 6.12±0.03 0.54 6.32±0.09 1.35
    下载: 导出CSV

    表 3  PANA 9600E系统抗干扰能力验证结果 log10IU/mL,X±S

    变量 胆红素(512 μmol/L) 血红蛋白(100 mg/mL) 甘油三酯(18 mmol/L)
    3.65 5.71 3.65 5.71 3.65 5.71
    对照组 3.66±0.05 5.71±0.09 3.66±0.05 5.71±0.09 3.66±0.05 5.71±0.09
    干扰组 3.61±0.01 5.74±0.06 3.60±0.03 5.73±0.01 3.66±0.02 5.67±0.08
    偏倚/% -1.38 0.49 -1.83 0.41 0.07 -0.67
    下载: 导出CSV
  • [1]

    World Health Organization(WHO). WHO global hepatitis report 2017[EB/OL]. [2021-07-18]. https://www.who.int/hepatitis/publications/global-hepatitis-report2017/en/.

    [2]

    Polaris Observatory Collaborators. Global prevalence, treatment, and prevention of hepatitis B virus infection in 2016: a modelling study[J]. Lancet Gastroenterol Hepatol, 2018, 3(6): 383-403. doi: 10.1016/S2468-1253(18)30056-6

    [3]

    中华医学会感染病学分会, 中华医学会肝病学分会. 慢性乙型肝炎防治指南(2019年版)[J]. 临床肝胆病杂志, 2019, 35(12): 2648-2669. https://www.cnki.com.cn/Article/CJFDTOTAL-LCGD201912010.htm

    [4]

    European Association for the Study of the Liver. Electronic address: easloffice@easloffice. eu, European Association for the Study of the Liver. EASL 2017 Clinical Practice Guidelines on the management of hepatitis B virus infection[J]. J Hepatol, 2017, 67(2): 370-398. doi: 10.1016/j.jhep.2017.03.021

    [5]

    Terrault NA, Lok A, McMahon BJ, et al. Update on prevention, diagnosis, and treatment of chronic hepatitis B: AASLD 2018 hepatitis B guidance[J]. Hepatology, 2018, 67(4): 1560-1599.

    [6]

    Cornberg M, Lok AS, Terrault NA, et al. Guidance for design and endpoints of clinical trials in chronic hepatitis B-Report from the 2019 EASL-AASLD HBV Treatment Endpoints Conference[J]. J Hepatol, 2020, 72(3): 539-557. doi: 10.1016/j.jhep.2019.11.003

    [7]

    肖圣达, 耿帜, 徐远东, 等. HBV DNA超敏定量检测试剂盒的性能验证[J]. 临床血液学杂志, 2021, 34(2): 81-85. https://www.cnki.com.cn/Article/CJFDTOTAL-LCXZ202102001.htm

    [8]

    Liu Y, Liu H, Hu Z, et al. Hepatitis B Virus Virions Produced Under Nucleos(t)ide Analogue Treatment Are Mainly Not Infectious Because of Irreversible DNA Chain Termination[J]. Hepatology, 2020, 71(2): 463-476. doi: 10.1002/hep.30844

  • 加载中

(3)

(3)

计量
  • 文章访问数:  412
  • PDF下载数:  297
  • 施引文献:  0
出版历程
收稿日期:  2021-08-04
修回日期:  2021-09-27
刊出日期:  2022-04-01

目录