电化学发光免疫法与ELISA法检验性能比较

韩骏飞, 姜宇, 宋发谷. 电化学发光免疫法与ELISA法检验性能比较[J]. 临床血液学杂志, 2019, 32(2): 152-154. doi: 10.13201/j.issn.1004-2806-b.2019.02.022
引用本文: 韩骏飞, 姜宇, 宋发谷. 电化学发光免疫法与ELISA法检验性能比较[J]. 临床血液学杂志, 2019, 32(2): 152-154. doi: 10.13201/j.issn.1004-2806-b.2019.02.022
HAN Junfei, JIANG Yu, SONG Fagu. Test performance comparison of HISCL-5000 and ELISA[J]. J Clin Hematol, 2019, 32(2): 152-154. doi: 10.13201/j.issn.1004-2806-b.2019.02.022
Citation: HAN Junfei, JIANG Yu, SONG Fagu. Test performance comparison of HISCL-5000 and ELISA[J]. J Clin Hematol, 2019, 32(2): 152-154. doi: 10.13201/j.issn.1004-2806-b.2019.02.022

电化学发光免疫法与ELISA法检验性能比较

详细信息
  • 中图分类号: R457.1

Test performance comparison of HISCL-5000 and ELISA

  • 目的:比较电化学发光免疫分析仪(HISCL-5000)与酶联免疫吸附试验(ELISA)的检验性能,为临床检验工作提供参考。方法:用HISCL-5000与ELISA法2种方法同时检测HISCL-5000配套的HBsAg校准品、质控品、临床样本以及2017年省免疫学HBsAg室间质评质控品,比较2种方法的线性范围、精密度和准确度。结果:HISCL-5000线性范围远大于ELISA法,HISCL-5000精密度和准确度均高于ELISA法,2种方法的差异有统计学意义(P<0.05)。结论:ELISA法线性范围较窄、精密度和准确度在手工操作过程中较差,易产生漏诊和误诊;而HISCL-5000具有高精密度、高准确度及良好线性范围的优点,可以替代ELISA法。
  • 加载中
  • [1]

    Wu CC,Chen YS,Cao L,et al.Hepatitis B virus infection: Defective surface antigen expression and pathogenesis[J].World J Gastroenterol,2018,24:3488-3499.

    [2]

    Akbar SM,Al-Mahtab M,Khan SI,et al. Current trends in hepatitis B vaccination[J].Future Virol,2016,11:369-378.

    [3]

    Liu C,Chen T,Lin J,et al.Evaluation of the performance of four methods for detection of hepatitis B surface antigen and their application for testing 116,455 specimens[J].J Virol Methods,2014,196:174-178.

    [4]

    Gish RG,Chang TT,Lai CL.Quantitative hepatitis B surface antigen analysis in hepatitis B e antigen-positive nucleoside-naive patients treated with entecavir[J].Antiviral therapy,2013,18:691-698.

    [5]

    Li L,Cai B,Tao C,et al.Performance Evaluation of CLIA for Treponema Pallidum Specific Antibodies Detection in Comparison with ELISA[J/OL].J Clin Lab Anal,2016,30:216-222.

    [6]

    Long A,Call DR,Cain KD.Comparison of quantitative PCR and ELISA for detection and quantification of Flavobacterium psychrophilum in salmonid broodstock[J].Dis Aquat Organ,2015,115:139-146.

    [7]

    Liu C,Chen T,Lin J,et al.Evaluation of the performance of four methods for detection of hepatitis B surface antigen and their application for testing 116455 specimens[J].J Virol Methods,2014,196:174-178.

    [8]

    Zhang X,Zhou D,Sheng S,et al.Electrochemical immunoassay for the cancer marker LMP-1(Epstein-Barr virus-derived latent membrane protein 1)using a glassy carbon electrode modified with Pd@Pt nanoparticles and a nanocomposite consisting of graphene sheets and MWCNTs[J].Microchimica Acta,2016,183:2055-2062.

    [9]

    Feng S,Wei B,Rao C,et al.Clinical Evaluation of the Newly Developed HISCL-5000 Analyzer on Detection of Hepatitis B Virus Markers in West China Hospital[J].Clin Lab,2016,62:1053-1060.

    [10]

    Ohne K,Kani S,Ohashi M,et al.Clinical evaluation of a newly developed high-sensitive detection of hepatitis B virus surface antigen by a semi-automated immune complex transfer chemiluminescent enzyme immunoassay[J].Rinsho Byori,2013,61:787-794.

    [11]

    Nakamura E,Kakuda H,Matsuura K,et al.Quantitative analysis of hepatitis B surface antigen as a clinical marker[J].Rinsho Byori,2011,59:838-843.

  • 加载中
计量
  • 文章访问数:  665
  • PDF下载数:  4261
  • 施引文献:  0
出版历程

目录