Verification before use of IEMS Microplate Incubator/Shaker
-
摘要: 目的:对新购入的THERMO IEMS微孔板震荡孵育器进行使用前确认,以保证使用该仪器得到的检测结果是正确有效的。方法:对已检测献血员样本54份及标准物质若干份进行检测并分析结果。结果:54份已检测献血员样本经FAME和孵育器两种方法检测结果一致。弱阳性标准物质经FAME和孵育器两种方法检测结果,经统计软件分析差异无统计学意义(均P>0.05),且精密度良好(批内CV<15%)。结论:THERMO IEMS微孔板震荡孵育器的结果与FAME相比完全吻合,可用于实验室的手工操作。Abstract: Objective: Verification before use of the newly bought THERMO IEMS Microplate Incubator/Shaker is necessary to ensure correct and effective tested result.Method: Test 54 samples from the blood donors already tested and standard materials and analyze the results.Result: The 54 tested samples from the blood donors are retested using FAME and incubator,the results are identical.The weak positive standard materials are also tested using FAME and incubator and analyzed by statistical software,the results are all P>0.05,showing no obvious difference and having good precision (RCV<15%).Conclusion: The test results using THERMO IEMS Microplate Incubator/Shaker coincides with that of FAME and it can be used in laboratory for manual operation.
-
Key words:
- incubator /
- verification before use /
- comparison experiment
-