新鲜全血在血液分析仪间比对及传递的临床应用

徐阳, 郑亚慧, 张磊, 等. 新鲜全血在血液分析仪间比对及传递的临床应用[J]. 临床血液学杂志, 2017, 30(10): 762-764. doi: 10.13201/j.issn.1004-2806-b.2017.10.007
引用本文: 徐阳, 郑亚慧, 张磊, 等. 新鲜全血在血液分析仪间比对及传递的临床应用[J]. 临床血液学杂志, 2017, 30(10): 762-764. doi: 10.13201/j.issn.1004-2806-b.2017.10.007
XU Yang, ZHENG Ya-hui, ZHANG Lei, et al. Clinical application of comparison of hematology analyzer by using fresh whole blood transfer method[J]. J Clin Hematol, 2017, 30(10): 762-764. doi: 10.13201/j.issn.1004-2806-b.2017.10.007
Citation: XU Yang, ZHENG Ya-hui, ZHANG Lei, et al. Clinical application of comparison of hematology analyzer by using fresh whole blood transfer method[J]. J Clin Hematol, 2017, 30(10): 762-764. doi: 10.13201/j.issn.1004-2806-b.2017.10.007

新鲜全血在血液分析仪间比对及传递的临床应用

详细信息
    通讯作者: 张磊,E-mail:zuibaxian@163.com
  • 中图分类号: R446.11

Clinical application of comparison of hematology analyzer by using fresh whole blood transfer method

More Information
  • 目的:探讨新鲜全血在血液分析仪比对及传递的临床应用。方法:用HST-302流水线1号机血细胞分析仪作为参比检测系统,对非血液系统疾病门诊患者新鲜全血白细胞(WBC)、红细胞(RBC)、血红蛋白(Hb)、血小板(PLT)和血细胞比容(HCT)进行定值,再用该血对HST-302流水线2号机进行比对校准,最后利用该定值新鲜全血在HST-302流水线1、2号机进行传递,Sysmex软件IPU有Cumulative累计功能,可保存检测结果,我们通过Cumulative即可随时监测仪器是否存在随机误差或系统误差。结果:通过新鲜全血对HST-302流水线1、2号机的比对,发现2台仪器一致性良好;新鲜全血传递可帮助我们及时发现仪器状态是否在控。结论:采用健康人新鲜全血比对校准后,提高了不同血液分析仪之间检测结果的一致性;同时利用新鲜全血在不同血液分析仪间进行传递是一种经济、实用和有效的措施,具有一定的临床可行性。
  • 加载中
  • [1]

    中华医学会检验分会血液学、体液学专家委员会.血液分析仪校准规范化的建议[M]//丛玉隆.血液学体液学检验与临床释疑.北京:人民军医出版社,2004:294-298.

    [2]

    庄俊华,黄宪章,翟培军.医学实验室质量体系[M].北京:人民卫生出版社,2006:217-217.

    [3]

    姚书克.尝试用新鲜全血替代血常规室内质控品的研究[J].国际检验医学杂志,2015,36(9):2555-2556.

  • 加载中
计量
  • 文章访问数:  228
  • PDF下载数:  195
  • 施引文献:  0
出版历程
收稿日期:  2016-12-26

目录