摘要: 目的:比较强生全自动玻璃珠血型及配血仪与手工微柱凝胶卡法在抗体筛选中的符合率和操作程序。方法:分别将标本用强生全自动仪器和手工微柱凝胶卡法进行检测,比较阳性标本的符合率,记录2种方法对不同标本的操作时间并统计操作程序。结果:2种方法的阳性结果符合率为100%,全自动方法较手工法检测时间较短。结论:全自动玻璃珠与手工微柱凝胶法检测不规则抗体完全符合,但全自动法比手工法操作更简单,更省时,错误率低。
-
关键词:
- 强生ORTHO AutoVue Innova全自动血型及配血仪 /
- 抗体筛选 /
- 操作程序
Abstract: Objective: To compare the coincidence rate and procedures in unexpected antibody screening between Johnson automated blood analysis system and the DiaMed manual micro-column gel typing system.Method: The samples were test by the Johnson AutoVue automatic devices and the method of manual micro-column gel typing system to compared the coincidence rate. Different number of samples were tested,using the two methods to calculate the time of operation and count the operation procedures.Result: The coincidence rate of positive samples of the two methods was 100%.The testing time of the manual micro-column gel typing system was longer than that of Johnson automated blood analysis system.Conclusion: The coincidence rate between the two methods was 100%,the operating procedures of the automated blood analysis system was easier and the error rate was lower. -
[1] 李哲.输血前不规则抗体筛选与输血安全[J].检验医学与临床,2012,9(6):720-721.
[2] 刘达庄.免疫血液学[M].上海:上海科学技术出版社,2002:187-193.
[3] 周金安,魏晴,涂同涛,等.微柱凝胶卡及试管法对抗D效价检测的比较[J].中国输血杂志,2008,21(3):197-197.
[4] 闫文强,曹奎杰,周金安.强生全自动血型配血仪与传统方法的应用比较[J].临床血液学杂志,2011,24(10):577-578.
[5] 王红,周金安,曹奎杰.全自动血型配血仪在血型检测中的应用[J].临床血液学杂志,2011,24(12):715-717.
[6] 王敏,王保龙,姚萍,等.全自动血型仪在输血前检查中的应用[J].临床输血与检验,2011,13(2):123-125.
计量